| Description | Line scan camera board with sensor without cover glass, with UV coating |
| Articlenumber (Eureca): | 02.29.0040 |
| Origin: | Germany |
| Manufacturer: | Eureca |
| EAN | 0675051081881 |
| Interface | USB |
| Sensor | TCD1304DG |
| Pixel number | 3648 |
| Pixel pitch | 8 µm |
| Pixel height | 200 µm |
| Minimum Exposure Time | 20 µs |
| Supply Current | < 250 mA |
| Supply Voltages | 5 V |
| Size (L×B×H) | 61 × 25 × 12 mm |
| Weight | 25 g |
| Remark | EDU (Educational) Version, with UV coating, without cover glass; available from Q2/2026 |
The e9u-LSMD-TCD1304-EDU was developed specifically for teaching purposes. Compared to the e9u-LSMD-TCD1304-STD, it offers deliberately reduced performance – but is significantly more budget-friendly and therefore perfect for internships, exercises, PoCs (proof of concept) and maker projects. It is aimed at professors, doctoral students, research assistants, and teachers, as well as tinkerers, makers, and pupils/students who want to work with real measurement data.
Hardware development has been completed and series production is being prepared (planned start: Q2/2026). We are currently polishing libraries, drivers, and documentation. Anyone who would like to get involved early on and provide feedback can already purchase prototypes as e9u-LSMD-TCD1304-EDU-EVAL at reduced prices. This EVAL variant is fully usable for teaching and laboratory environments, but is not yet suitable for safety-critical applications.
| Operation mode | 8-bit mode | 12-bit mode |
|---|---|---|
| Digitization Depth | 8 bit | 16 bit left aligned |
| Read out noise (typ. rms) | 300 e− | 50 e− |
| Linear Full Well Capacity | 40 ke− | 40 ke− |
| Full Scale Capacity | 65 ke− | 65 ke− |
| Conversion Gain (typ @ 0.4 FS) | @ 8 Bit: 330±100 e−/LSB | @ 16 Bit: 1.3±0.4 e−/LSB |
| Maximum Frame Rate | 25 fps | 25 fps |
| Integration Start/Stop via USB | ✗ | ✗ |
| Start/Stop via USB Latency | 10 ms | 10 ms |
| External Trigger Input | ✓ | ✓ |
| External Trigger Latency | 20 µs | 20 µs |
| External Trigger Jitter | ±10 µs | ±10 µs |
| External Trigger Latency (debounced) | ✗ | ✗ |
| Isolated external Trigger Input | ✗ | ✗ |
| External Exposure Time Control | ✗ | ✗ |
| Exposure Time Control Latency | ✗ | ✗ |
| Minimum External Exposure Time | ✗ | ✗ |
| Maximum External Exposure Time | ✗ | ✗ |
| Exposure Time Control Latency (debounced) | ✗ | ✗ |
| Minimum External Exposure Time (debounced) | ✗ | ✗ |
| Trigger Output | ✓ | ✓ |
| High Resolution Timestamp: Trigger | ✗ | ✗ |
| GPIO | ✗ | ✗ |
| Enhanced Frame Buffer | ✗ | ✗ |
This module uses the TCD1304DG sensor without the standard cover glass and with an additional Lumogen coating to extend the usable spectral range into the UV. This makes the module ideal for use in spectrometers for e. g. the material analysis.
The cover glass is removed and the coating applied in-house at Eureca using specially developed processes. After removing the cover glass, all sensors are tested for their function on the respective camera module and taped with a protective film for transport. This must be removed before using the module. If you need permanent protection, we recommend a cover with quartz glass (TCD1304DG-Q).
Since the sensor is mounted on the module, it can easily be replaced by another sensor version if required. This can e. g. be the sensors in the standard version. But sensors only without a cover glass or sensors with special filters are also possible here.
After replacing the sensor, the module must be recalibrated. A special automatic program is available for this, which is included in the scope of delivery of the modules.
For some of these variants, there are already module versions with the corresponding preinstalled modified sensors. You will find a corresponding list at the end of this page.
Further information, source codes and documentation can be found on our service page!
For orders (also from end customers) please use the link below.
For offers for larger quantities, please contact us by e-mail.